Alignment of the KappaCCD2000 on an FR591 rotating anode generator.

General principle

The alignment of the KappaCCD2000 on an FR591 rotating anode uses an alignment stage. This alignment stage is designed to make uncoupled translations and rotations of the diffractometer, to make the alignment of the diffractometer a fast converging process.

Alignment of the goniometer consists of two translations and two rotations perpendicular to the beam, one up and one horizontal, parallel to the table. The translation up is done with two height adjustments at the side of the goniometer. The rotation up is done with a single height adjustment at the end of the goniometer, below the theta arm.

The two horizontal motions (rotation & translation) are done with two micrometers, which we call Micrometer #1 and #2. These micrometers move the goniometer relative to a metal bar, which must be firmly fixed to the generator tabletop before using the micrometers. During the use of the micrometers, the hover pads must be activated.

There are two versions of the alignment plate. In the first version, the hover support positions and the diffractometer feet positions, do not coincide. Here the micrometers are positioned at two different sides of the goniometer. This is shown in the pictures below.
Alignment table 1 top view
Photo of alignment table

In the second version of the alignment table, the hover support positions and the diffractometer feet positions coincide. Also the micrometers are positioned adjacent to one another at the back of the goniometer. This is shown in the drawing below:
Alignment table #2

Alignment procedure

The alignment procedure of the goniometer assumes that the collimator is optimally aligned and points at the centre of the goniometer. When we are sure of this, we take the following steps: Alignment of the multilayer graded mirror is done using a Xyclops X-ray imager.

Xyclops X-ray imager This small camera can be connected to the multilayer graded mirror housing using a Xyclops adapter. The Xyclops and the mirror adapter are essential to make sure the X-ray beam shoots through the center of the mirror output port. This is crucial for getting an optimum intensity and alignment of the goniometer.

To make sure the beam is centered we use two alignment collimators:

Alignment collimators side view alignment collimators front view

One is completely open and is called the open-beam pipe. The other one is screened with a half-moon at the entrance. This is called the half-beam pipe. The orientation of the half moon is indicated with dots at the outside of the collimator.

o facing up Half moon up
oo facing up Half moon left
ooo facing up Half moon down
oooo facing up Half moon right

The alignment collimators are always used to look at the primary beam falling on the detector and therefore they should always be used at relatively low power (45kV, 35mA) together with the absorber.


Half beam absorber bar

For centering the beam at the crystal position, a 1.5 mm bar on a goniometer is used.

Half beam absorber bar

When properly centered, this should absorb the beam half-way; hence it is called the half-beam absorber. This 1.5 mm bar must be aligned using the video microscope and the crystal alignment routine on the server PC. Use position / (former view 1)and use the button Rotate 180°: the edge of the bar must be in the centre (it must absorb half of the beam). But it is better to leave a gap then to have overlap between the two positions. You are advised to keep the half-beam absorber aligned on a goniometer head for future alignments.

Coarse adjustment

To get close to the position where the goniometer is more or less at the same height as the direct beam we use a tool which is put on the collimator and which, if the goniometer is at the right height, blocks the direct beam in half. This is the "half absorber yoke".

Half beam absorber yoke

At this point you can see the direct beam through the largest collimator available (the open beam pipe). From here on we go to the fine alignment, which is also where you would normally start after changing a filament.

Fine adjustment

Before starting on this section makes sure that:

Take care: When changing the open beam pipe, always make sure the shutter is closed by checking the shutter open lamp!

Initial run:

Final run:

For normal measurements there are three standard collimators.
Measuring collimators Put in the largest (normal measurement) collimator with the attenuator and measure the intensity under similar circumstances used previously. Do the same with the other collimators. Write these values down for future reference.

This finishes the fine alignment of the goniometer.

Mounting and dismounting the multilayer graded mirror

Dismounting

Remounting

First installation of the multilayer graded mirrors


NoniusTo the Bruker Nonius Home

(C) Bruker Nonius BV 2002. Last change 15 March 2002