Cell determination using Dirax
The "cell determinatiuon using Dirax" is an alternative to the "cell
determination using Denzo". It uses a new method of determining the
cell dimensions. To be able to use this method, the detector
calibration needs to be known (please see manuals for makedistor and makedetalign) and the dirax program needs
to be installed (see variable dirax_installed in the configuration manual).
This new method of cell determination using an area detector and a
full 4-circle goniostat is very powerful in discriminating multiple
crystal lattices and/or indexing incommensurate unit cells.
The experiment
The experiment consists of 3 buttons:
- Make Scan-set: design the measurement.
- collect data: perform the measurement.
- Find cell using PhiChi: run the phichi program to find the unit cell from the measured frames.
Click on "Find cell using PhiChi". A scan set window appears.
Click on "Strategy."
The four parameters are:
- The detector to crystal distance (dx)
-
A crystal to detector distance of 40 should be OK
for cells with axis lengths upto about 30 A. If there are longer
axes, it will be difficult to separate the spots. For determining
the unit cell it is more important to be able to separate all
spots than it is to have many spots, hence it is not so bad to be
a bit further away than strictly necessary. The default can be
changed in the configuration parameters.
- Number of scan pairs
-
In principle the phi/chi indexing method requires only one scan pair
(one phi and one phi/chi) scan to determine the unit cell. However, if
the measurement is repeated 180 degrees further in phi, all
reflections are measured in Friedel pairs. This averages out some
systematic errors, and tends to give better results. Furthermore, all
reflections measured in one set come from a single region in the Ewald
sphere probably making one of the unit cell parameters less accurate
than possible. It is thus wise to make another 2 pairs at phi=90
degrees and phi=270 degrees. This makes up the default 4 pairs. To
increase the number of reflections found, this number can be raised to
higher even integers.
- Phi scan angle
-
This specifies how far to rotate phi in a single phi/chi scan pair.
This should be chosen such that a significant number of reflections
occur in the frame, without much overlap.
- Chi scan angle
-
This specifies how far to rotate chi in a single phi/chi scan.
This should normally be set to 10 degrees, independent of the
Phi scan angle.
Clicking on "More options" will bring up 3 additional parameters:
- Total phi range covered
-
Should normally be 360 degrees. This makes the coverage of the ewald
sphere homogeneous.
- First phi value
-
This is normally -180 degrees. If you do not have enough data, you can set
this parameter to -135 and measure an additional 4 scan pairs.
- Detector theta
-
This is normally 0 degrees. If there are very little reflections because
of an extremely small unit cell, it might be necessary to perform a phi/chi
experiment with a detector theta offset.
Click "OK". A window will appear asking for the KappaCCD server to connect
to. Complete its questions at click "OK". Almost instantaneous, the
complete phi/chi strategy will pop up.
You can add other operations to change cryostat and generator settings,
but you should not change the order of the scans.
Click "OK" to continue. A scan parameter window will pop up.
Unless you already know that the crystal diffracts either very well or
very bad, leave the suggested integration time. Under "More options"
some other standard parameters can be
changed. Click "OK" to continue.
The measurement will now be performed, and the unit cell will be
calculated. The resulting unit cell will pop up on the screen in a
window offering you the option to run the ndirax program to
play with the peaks that were found.
This completes the phi/chi experiment.
What can you do to improve the accuracy?
- You need to make sure that the detector settings are very accurately
known. Run makedetalign on a
well-aligned ammonium bitartrate test crystal first.
- You can use longer measurement times, improving the accuracy with which
reflections can be located.
- If you want to spend even more time, you can make more than 4
scans (under "More options" in the strategy parameters).
The standard scan parameters are pretty good guesses for normal unit cell sizes. But:
- If you do not get enough reflections: increase phi range to get more.
- If you have high mosaicity: increase phi range to make sure there
are enough complete (as opposed to partial) reflections in the images.
Increasing the chi range can improve the accuracy of the reflection
positions, but it can cause confusion between reflections if there are
many. Furthermore, very long chi scans (>20 degrees) can not be
accurately performed.