Bruker AXS index previous next

Previous: Cell determination using Denzo
Next: Orientations and single scans


Cell determination using Dirax

The "cell determinatiuon using Dirax" is an alternative to the "cell determination using Denzo". It uses a new method of determining the cell dimensions. To be able to use this method, the detector calibration needs to be known (please see manuals for makedistor and makedetalign) and the dirax program needs to be installed (see variable dirax_installed in the configuration manual).

This new method of cell determination using an area detector and a full 4-circle goniostat is very powerful in discriminating multiple crystal lattices and/or indexing incommensurate unit cells.

The experiment

Collect: phi chi experiment

The experiment consists of 3 buttons:

Click on "Find cell using PhiChi". A scan set window appears.

Empty scan set window

Click on "Strategy."

Strategy parameters

The four parameters are:

The detector to crystal distance (dx)
A crystal to detector distance of 40 should be OK for cells with axis lengths upto about 30 A. If there are longer axes, it will be difficult to separate the spots. For determining the unit cell it is more important to be able to separate all spots than it is to have many spots, hence it is not so bad to be a bit further away than strictly necessary. The default can be changed in the configuration parameters.
Number of scan pairs
In principle the phi/chi indexing method requires only one scan pair (one phi and one phi/chi) scan to determine the unit cell. However, if the measurement is repeated 180 degrees further in phi, all reflections are measured in Friedel pairs. This averages out some systematic errors, and tends to give better results. Furthermore, all reflections measured in one set come from a single region in the Ewald sphere probably making one of the unit cell parameters less accurate than possible. It is thus wise to make another 2 pairs at phi=90 degrees and phi=270 degrees. This makes up the default 4 pairs. To increase the number of reflections found, this number can be raised to higher even integers.
Phi scan angle
This specifies how far to rotate phi in a single phi/chi scan pair. This should be chosen such that a significant number of reflections occur in the frame, without much overlap.
Chi scan angle
This specifies how far to rotate chi in a single phi/chi scan. This should normally be set to 10 degrees, independent of the Phi scan angle.
Clicking on "More options" will bring up 3 additional parameters:

Strategy parameters; expert view

Total phi range covered
Should normally be 360 degrees. This makes the coverage of the ewald sphere homogeneous.
First phi value
This is normally -180 degrees. If you do not have enough data, you can set this parameter to -135 and measure an additional 4 scan pairs.
Detector theta
This is normally 0 degrees. If there are very little reflections because of an extremely small unit cell, it might be necessary to perform a phi/chi experiment with a detector theta offset.
Click "OK". A window will appear asking for the KappaCCD server to connect to. Complete its questions at click "OK". Almost instantaneous, the complete phi/chi strategy will pop up.

Completed Phi/Chi strategy

You can add other operations to change cryostat and generator settings, but you should not change the order of the scans.

Click "OK" to continue. A scan parameter window will pop up.

Phi/Chi scan parameters

Unless you already know that the crystal diffracts either very well or very bad, leave the suggested integration time. Under "More options" some other standard parameters can be changed. Click "OK" to continue.

The measurement will now be performed, and the unit cell will be calculated. The resulting unit cell will pop up on the screen in a window offering you the option to run the ndirax program to play with the peaks that were found.

This completes the phi/chi experiment.

What can you do to improve the accuracy?

The standard scan parameters are pretty good guesses for normal unit cell sizes. But: Increasing the chi range can improve the accuracy of the reflection positions, but it can cause confusion between reflections if there are many. Furthermore, very long chi scans (>20 degrees) can not be accurately performed.

index previous next

Previous: Cell determination using Denzo
Next: Orientations and single scans


(C) 1997-2009, Bruker AXS BV, R.W.W. Hooft