Nonius logo KappaCCD User Manual 
 Revision 1.10, Last Update 1-July-1998

Calibration

Calibration comes in several flavors. Calibratable items are for instance distortion, sensitivity, camera orientation and position, and dark current. Of all these calibrations, only camera orientation and position can be calibrated on site and should be repeated when the X-ray tube is changed or other alignment work is done.
 

Calibration procedure after alignment.

Collect a standard data set on the test crystal.
 
Data collection parameters 

Position: 
Theta=0 
Phi=0 
Omega=160 
Kappa=0 
Dx=25 

Scan:  
25 seconds 
90 frames 
repeat=1, delta phi=2 

Do NOT use the factory dataset provided with the system!  
Calibration has to be done with a locally collected data set in order to describe camera settings after final installation.
   
    Refinement window output  
     
    • Cassette RotX
    • Cassette RotY
    • (Cassette RotZ remains zero)
    • Xbeam
    • Ybeam
    • Crossfire X
    • Crossfire Y
    • Crossfire XY
    User Interface Site Card  
     
    • Rotation X
    • Rotation Y
    • Rotation Z (=0)
    • Beam Position X
    • Beam Position Y
    • Crossfire X
    • Crossfire Y
    • Crossfire XY
 
 
 

Test the calibration

You can test the calibration by loading the new Initial.set file and index again the same data set. Successful calibration will yield very low Figures of Merit in your indexing table.