Previous: Data Collection
Next: Alignment of the diffractometer
NONIUS CAD4/MACH3
21 December 2000
    Quick overview of CAD4 Commands

This section gives a quick overview of CAD4 Commands. For a more detailed description of commands you are referred to the CAD4 User manual. The links of the command point to specific sections in the On-line CAD4 user manual. The reference between brackets point to the page of the hardcopy printed CAD4 user manual.

Basic Diffractometer Control and Specifications
Command (Ref.) Action
GONCON (7-1) Modify GONiometer CONstants and file assignments
RA (8-1) Remove Attenuator
RB (8-1) Remove Balanced filter
SA (8-1) Set Attenuator
SAP (8-2) Set horizontal Aperture
SBA (8-1) Set Bal. Filter A (Not in newer systems)
SBB (8-1) Set Bal. Filter B (not in newer system)
SC (8-1) Shutter Close
SH(8-2) Set Horizontal slit
SL(8-2) Set Lower half-circle aperture
SN (8-2) Set Negative skew slit
SO (8-1) Shutter Open
SP (8-2) Set Positive skew slit
SU (8-2) Set Upper half-circle aperture
SV (8-2) Set Vertical slit
WI (8-8) Wavelength(s) Input, Attenuation factor
WO (8-8) Wavelength(s) Output, Attenuation factor
Goniometer positioning
Geometry transformation commands (Ref: 8-2,3,4)
HH HC HB HE HK HP H = Miller indices
CH CC CB CE CK Cp C = Cart coords of refl.
BH BC BB BE BK BP B = Bisect_’s (?=0)
EH EC EB EE EK EP E = Eulerian_’s (??0)
KH KC KB KE KK KP K = Kappa angles
MH MC MB ME MK MP M = Measure
.H .C .B .E .K .P . = "geometry buffer"
PP (=.P) P = Position the angles
.O Output result of last calculation
Related Positioning Commands
Command (ref) Action
AA(8-5) Alternative Angles setting of current refl
NH(8-5) Set "inverse hkl"reflection
NN(8-5) Set inverse hkl at inverse 2-theta
TN(8-5) Set inverse 2-theta position
Special Positions
Command(Ref) Action
DEMO(8-6) Goniometer demonstration
MICROR(8-7) Index and measure crystal faces
MICROS(8-7) Set a face normal in beam direction
VIEW1,2.. (8-6) Microscope viewing positions
ZERO(8-6) Set all angles to zero positions
Reflection Locating, Centering and Scanning
Command (Ref) Action
AXIAL(10-58) Make axial oscillation (omega axis) photographs
DETTH(10-8) Determine 2-theta angles from +/- measurements
PHOTO(10-5) Find and center Refls from a rotating crystal photograph
POLA(8-6) Polaroid rotation photo
SCAN(10-29) Goniometer scan through present position
SEARCH(10-2) Search for reflections (spiral or shell)
SET4(10-10) Set angles free from zero and xtal XYZ errors
SETANG(10-9) Set angles of Center reflections from the list
SETPAR(8-7) Set parameters for reflection centering
List Operations
Command(Ref.) Action
LD(9-11) List dump (transfer from "geometry buffer" into list
LG(9-11) List Get- Get refl from the list (transfer into "geometry buffer
LH(9-8) Input indices (known orientation)
LI(9-7) List input (indices and angles)
LK (9-7) List Kill – Delete reflections in the list
LO(9-2) List Out – Output list entries
LPH(9-8) Input x,y coordinates from Polaroid photo
Status Code Modification
Command (Ref.) Action
LCA(9-12) Change status code of Angles (for SETANG)
LCH(9-11) Change status code of Indices (for INDEX)
LCI(9-13) Change status code for Intensity monitoring
LCO(9-12) Change status code for Orientation monitoring
LCS (9-13) Change status code for Scan information
LCT(9-12) Change status code of Theta (for DETTH)
Orientation Matrix Determination and Indexing
Command (Ref) Action
CELDIM(10-30) Cell-dimensions LS using theta values only, constraints
INDEX(10-17) Auto-indexing procedure with LS
LS(10-28) Least-squares refnmnt matrix elements. Extract cell params
RAMCEL (10-12) Matrix from cell + 2 reflections
REIND (10-21) Re-Index with present matrix with LS
RI (8-7) Input orientation matrix
RO (8-7) Output orientation matrix
TRANS (10-22) Cell reduction & transformation. Conventional cell from metric
Data Collection
Command (Ref.) Action
DATCIN (11-3) Data Collection parameters Input
DATCOL (11-20) Start data collection
DATCON (11-21) Data collection Continue
DATHKL (11-21) Prepare list of reflns to be collected
DATAR (11-39) Read and display selected reflections from data file
TH (10-29) Number of reflections and index limits in theta range
Diffractometer Alignment
Command (Ref.) Action
ALIGN (12-13) Align diffractometer offsets from 8-reflection centering
DEADT (13-5) Counter DEAD Time and attenuation factor
Switch Register Codes (select output options)
Setting(Ref.) Action
4XXX (6-13) Results from DATCOL / Repeat of stationary SCAN
2XXX (6-13) Profiles from SEARCH, PHOTO, SETANG, DETTH
1XXX (6-14) Integral and frActional indices DATCOL, orientation
X4XX (6-15) Bgd, peak, descriminator value from SEARCH, PHOTO
X2XX (6-16) FrActional indices, angular deviation from LS, TRANS
X1XX (6-16) To supply cell-dimension, INDEX
XXX1 (6-1) SWITCH 11 interrupt (pocket terminal)
To obtain combination, set the switch register to the sum of the appropriate settings.
Miscellaneous
Command(Ref.) Action
CONSW (1-10) Switches to terminal on the CT port (PDP-11 and VAX only)
/ (6-11) The original value of a parameter is retained
<CTRL>C (6-12) Interrupts a CAD-4 program. A"/" prompts "GO AGAIN?"
Q Terminate input and return to the command mode (CD4>)
EXIT Exit CAD4
DATCIN Parameters
Name (Ref) Parameters Meaning
TEXT (11-6) File Ident
THLIM (11-6) MIN
MAX
Min theta values
Max theta values
SCAN (11-7) DOMA, DOMB
APTA, APTB
TYPE
Omega scan angles
Horiz. width of detector
Type of prescan and final scan
INT (11-8) SIGPRE
SIGMA
NPIPRE
MAXTIME
DUMPFL
Prescan accept. Parameter
Sigma(I)/I for final scan
Prescan speed parameter
Max time for final scan (sec)
Intensity profile output
FLAG (11-9) NBAFL
NFRIDEL
NEQFL
Balanced filter
Measurement of friedel pairs
Non-equal test
PSI (11-11) WORD

N
PSIST
DELPSI
V1, V2, V3

Controls the azimuth angle
(BISECT, AZIMUT, FLAT, NEEDLE)
Number of final scans (NRPSI)
Starting psi value
Increment in degrees
Vector normal to FLAT crystal or in NEEDLE
MODE (11-13) WORD

HMIN, HMAX….
HST, KST, LST
SEQ

Seq mode for data collection
ZIGZAG, FILE, SEPHKL
Hkl range
Hkl starting values
Incremental order (HKL, HKL,…fast.med,slow.)
ABSENT (11-14) NR
XN, YN, ZN
Number of absence conditions
Absence conditions
INTCR (11-15) REPTIME
SIGMI
FRACTION
Exposure time between intensity monitors
Sigma(I)/I ratio of intensity monitor
Intensity monitor cut-off (FADING)
ORIENT (11-17) REPCNT
DEV-ANG
Refl count between orientation monitors
Max allowable angular deviation
BACKG (11-19) BGSEL
BGTIME
BGDIST
Static background parameter
Static background in seconds
Relative position parameter for static background
Typical Values for DATCIN
Molybdenum Radiation Copper Radiation
TEXT Name, date, mA, kV Name, date, mA, kV
THLIM 1 25 1 55
SCAN 0.80 0.347 2.70 0.40 6 0.80 0.142 2.70 0.40 6
INT 2.0 0.03 3 60 0 2.0 0.03 3 60 0
FLAG 0 0 0 0 0 0
PSI BISECT
37 0 10 1 0 0
BISECT
37 0 10 1 0 0
MODE ZIGZAG
-5 5 –5 5 –5 5
-5 –5 –5
LKH
ZIGZAG
-5 5 –5 5 –5 5
-5 –5 –5
LKH
ABSENT 0 0
INTCR 3600 0.01 0.10 3600 0.01 0.10
ORIENT 400 0.15 400 0.15
BACKG 0 2 1 0 2 1


Previous: Data Collection
Next: Alignment of the diffractometer
(C) Nonius BV, 1999, 2000. Last change 21 December 2000