| FR590 |
 |
July 2009
0590.761
Introduction
The FR 590 is a high voltage supply for a sealed X-ray tube and consists
of:
-
High voltage generator.
-
Safety circuit, when connected to a Nonius goniometer
-
Cooling water connections with
-
Cock to empty the tube's water circuit.
-
Water flow switch for check on minimal water flow. There is no check on
the input water temperature.
Safety
instructions
Alterations, other than mentioned in this manual, can obstruct the safety
precautions and are therefore not allowed. Parts may only be changed by
identical parts.
X-ray
safety
-
The equipment must be operated under the responsibility of an X-ray safety
officer, as stated by local regulations.
-
The equipment must be operated, serviced and maintained by qualified personnel
who are trained with the hazards of and precaution against X-ray exposure.
-
The equipment contains safety circuits which should not be overridden,
-
The X-ray key should not be kept with the instrument, but under the custody
of the X-ray safety officer in case X-ray is not allowed to be used.
-
Safety circuits must be tested by service personnel or the safety officer
at least every 12 months, as described in the manual of the goniometer.
Beryllium
The X-ray tube contain windows of beryllium which can be corroded. This
corrosion is very poisonous. When removing the tube these can be reached
and care must be taken not to touch these window. When by accident the
beryllium is touched, wash with soap and water. See also the general
explanation of Be hazards or the
X-ray
tube mounting instructions. From mid 2000 the X-ray tubes will be delivered
with a protective cover :
-
Remove bag and cover only just before installation in the system.
-
When the tube is removed from the system for a short time, the cover should
be put back on the tube.
-
When removed from the system for a longer time or permanently the tube,
with protective cover, should be packed in a plastic bag (after removing
all cooling water). A spare bag is added with the tube
Fumes and dust from beryllium metal can be hazardous if inhaled. During
use, corrosion of the beryllium window can occur, but it should not be
cut, machined or otherwise removed. Disposal must comply with the national
and local regulations governing beryllium.
Electrical safety
Access to the area inside the generator is only allowed by qualified
technical personnel. Open power terminals can be reached inside the generator.
When a qualified technician wants to access these parts or the HV cable,
first disconnect the power connector (don't touch the connector pins) and
wait several minutes to allow discharge of all components.
Specifications
| X-ray |
|
| Voltage |
20 - 60 kV in 1 kV steps |
| HV ripple |
0.7 V rms. at full load |
| Current |
5 - 60 mA in 1 mA steps |
| Stability |
Voltage and current : 0.01% |
| Power (kW) |
max. 3 kW, depending on the tube |
|
|
The X-ray high voltage is not switched on, or is switched off when in the
Nonius goniometer:
-
The shutter does not close.
-
The high voltage cable is not connected to the X-ray tube (indicated by
a link, attached to the cable)
-
The X-ray on lamp is defect.
Environmental, according to IEC 1010
| Environment |
Indoor use |
| Altitude |
Up to 2000 m |
| Temperature, operating |
max. 40 °C |
| Relative humidity |
Maximum 80% if <31°C, decrease linearly to 50% at 40°C |
| Pollution degree |
2, according to IEC 664 (normal clean environment). |
| Protection |
IP 20, according to IEC 529 (limited protection to water). |
| Gasses |
No explosive gasses are allowed. |
| HV tank oil |
Shell Diala D. The oil does not contain PCB's. This oil is also
used in FR571, GX21 and FR559.. generators |
|
|
| Power supply |
Nominal 200, 208, 220, 230 or 240 V, single phase |
| Fluctuations |
+10%, -12% |
| Frequency |
50 - 60 Hz ±5% |
| Mains fuse |
35 A |
| Installation category |
II, according to IEC 664. |
|
|
| Water cooling |
|
| Flow |
For optimal tube life time: stable 4 - 5 l/min,
Minimum: 3.5 l/min, max: 6 l/min |
| Temperature |
15 - 25 ºC depending on humidity. General advise : not lower then
ambient temperature to prevent condensation. |
| Water
for maximum tube life |
Normal tap water quality can be used or better: demineralized water
with :
- specific conductivity: 10 - 250 uS/cm
- hardness : 1 - 4 D
- pH : 7.5 - 9.5
- no contact with air
- no additives |
| Input pressure |
4 - 6 Bar |
|
|
| Weight |
80 kg |
Operation
X-ray on
-
Switch the key (power) on.
-
The green "system OK" lamp on the front panel will lit.
-
Press: X-ray on.
-
The X-rays on lamp on X-ray tube and the front panel will burn..
-
The power relay switches on (audible).
-
After approx. 5 s first kV and then mA will rise to the chosen value (minimum
20 kV 5 mA).
Remote control
-
To go to remote control (so an external program like KappaCCD or MACH3
can control the kV/mA):
-
Press 590 <return> 9 <return> <return> (<return> mains while
keep pressing 1 press <enter>)
-
After the second <return> you will get : ........for a figure
code (password). Normally don't use this.
-
Leaving remote control :
-
Press <return>, 2x <CR>
-
When using a figure code (password): code, <return>, 2x <CR>
Special commands
With special commands in the "590 mode", internal data can be read:
-
Go into 590 mode with: 590 <enter>
-
Choose the code with: code number <enter>
-
Modify : keep the last number of the code number pressed during <enter>.
-
When pressing <enter> the second number of the code will go up.
-
One step back with: <CR>
-
Go out of the 590 mode with: several <CR>
Codes (590 <enter> code number), a selection:
| Code number |
Also modify |
|
|
| 10 |
X |
hh:mm:ss |
Time |
| 11 |
X |
dummy |
Date |
| 20 |
|
UFO = ##. |
real kV = "output" front display |
| 21 |
|
Io = ##. |
real mA = "output" front display |
| 22 |
|
OIL = ##. |
Oil temperature in degrees |
| 23 |
|
CAB = ##. |
Inside air temperature in degrees |
| 24 |
|
UPI = ##. |
setpoint kV |
| 25 |
|
Ii = ##. |
setpoint mA |
| 26 |
|
FILL = ##. |
Filament voltage |
| 27 |
|
UDC = ### |
UDC voltage |
| 3n |
X |
|
Tube parameters of tube n :
TUBE_n : Tube number
Im = ###mA : Maximum current, default 60 mA
Um = ###kV : Maximum voltage, default 60 kV
Pm = ##.#kW : Maximum power, depending on tube (max. 3 kW)
T = #####h : X-ray on time in hours, |
| 4nn <enter> |
|
|
Displays the error history, n = 0 = last,
max. 53. Displays error code, date and time |
| 50 |
|
TS = ##### |
Total power on time in hours |
| 51 |
|
TX= ##### |
Total X-ray on time in hours |
| 70 |
|
|
Battery life left, calculated in % of expected life |
| 91 |
X |
|
Remote. Switching to remote. Displays the baud
rate. |
Maintenance
Periodic maintenance
A yearly check is advised, which contains:
-
Check of the water flow, the water flow switch and the water temperature.
Because the flow switch can be polluted, depending on water quality, it
is advised to check this more frequently, like ones per month. This quick
and rough test can be done by switching off the water cooling and then
switch on the X-ray shortly, with the lowest setting 20 kV 5 mA.
-
Clean the internal ventilator, when it is very noisy, it should be replaced.
-
When combined with a goniometer with tube: Open the X-ray tube and clean
the water filter and check the X-ray tube on overheating.
-
Check the battery life Zero power RAM. When it is below 15%, it is advisable
to replace.
-
Check the safety with the goniometer safety check.
The check can be done by Service personnel, combined with the goniometer.
The items of the FR590 can also be done by a qualified technician of the
customer.
X-ray tube
See the instruction manual of the X-ray tube, which is delivered with
every new tube (at the bottom of the box). A drawing is also in the last
pages or on page 42 of the old 590 User manual. Some items:
-
For long life time :
-
Don't switch off from full power : first lower the mA, then the kV, then
switch off.
-
For power up: first set the kV, then the mA.
-
When the tube is new or not used more then 24 hours a conditioning procedure
is advised (see the Tube manual).
-
When not used for a short time : leave the tube on 20 kV, 5 mA (Siemens
tube) or 40kV 10 mA (Philips tube) instead of switching off.
-
Regularly feel by hand the temperature of the tube. When it gets warm,
clean the tube inside
-
Clean the inside of the tube at least every
1000 hours of operation.
-
Take into account there are long (B) and short (A) tubes. Long tubes needs
an adapter. Nowadays short tubes are delivered, for which no adapter is
used. Don't use a long tube without the adapter, because this will destroy
the tube.
Mounting of a X-ray tube
-
Dismounting :
-
Turn off the mains.
-
Close the external inlet valve of the water supply (as an extra precaution).
-
Turn the drain valve to the drain position.
-
Connect the bicycle pump and pump until you notice a decrease in pressure.
This will empty the tube's water circuit as far as possible.
-
Remove the X-ray tube (don't remove the HV cable).
-
Be careful: the X-ray tubes has beryllium windows which could cause safety
hazards. See safety instructions
-
Handled the tube by holding the cooling plate or by the cylindrical metal
part between the glass/ceramic insulator and the window area. Do not handle
the tube with the new safety cover, because it may slip away from it
-
Take into account that not all water is removed.
-
Check for leaks near the X-ray tube.
-
Open and clean the tube
-
Mounting :
-
Make sure that the interior of the tube shield is dry and clean.
-
With a new tube with safety cover : remove also the elastic ring
-
Mount the tube
-
Open the outside water inlet valve.
-
Turn the drain valve to the cooling position.
-
Turn the mains on. The system has noticed that a tube is changed and starts
asking for the tube parameters.
-
In case of a used tube: Select the proper tube number and press <enter>
for the other parameters.
-
In case of a new tube:
-
Choose a new tube number (or just the default 0)
-
Input the maximum tube voltage, mostly the default 60 kV with <enter>
-
Input the maximum tube current, depending on the tube used (see the Tube
manual: fine focus : 40 mA)
-
Input the maximum tube power, depending on the tube used.
-
Input the number of hours the tube has already been used with <enter>
-
If you made a mistake press <CR>.
-
If you have made a mistake after all date is entered, just switch the drain
valve on and off and start again.
-
It is possible to check if the right values are set with the special
commands
-
Initialize the new tube as described in the tube manual
-
Siemens ceramic tubes and KappaCCD. The Collect program has a command:
newtube [scenario] with executes this startup automatically. See the manual
of Collect for a description.
Cleaning of a X-ray tube
See the Tube instruction manual delivered with tube :
-
Remove the X-ray tube
-
Remove the four screws in the top cap.
-

-
Remove the water nozzle but do not deform it (do not use pliers when the
nozzle is stuck). Clean it, using methanol, a lint free cloth or Scotch
Brit (type 7558).
-
Remove the filter and clean it.
-
When there is deposit on the anode cooling face (the gold pyramid structure)
:
-
Put the tube with the anode up.
-
Fill the anode coupling space with acetic acid (10%).
-
Scrub the pyramid with a small, soft brush with short hairs.
-
Leave it for 20 minutes and scrub again.
-
Flush the area with water and refill with acid.
-
Repeat until the anode is clean.
-
Fit the water filter.
-
Mount the nozzle till the position hole slides just over the positioning
pin. Do no push it completely till the base plate, because during final
mounting the nozzle must be pressed against the cooling surface of the
anode.
-
Place the tube on the base plate, ensure that the sealing rings are in
place and tighten the four screws. One of these screws is positioned asymmetrically
to ensure correct positioning.
Disposal
-
The HV tank contains oil without PCB's. A statement in Acrobat .pdf form
can be retrieved
-
The X-ray tubes contain beryllium windows. Disposal must comply with the
national and local regulations governing beryllium.
Parts
Zero power RAM
Calibration values are saved in a memory with an internal battery :
the Zero power RAM. When the battery is empty the values are lost and the
error UNCALIBRATED appears. The battery live is about 3 years of non power
buffering + time the generator is on, total mostly about 7 years. The software
is calculating (not measuring) the life time which can be read
out. It is advisable to change the Zero power RAM when it reaches 0%,
although a longer life is possible. All the tube data and total timing
data will be lost. When the battery is empty the generator will remain
working, but after every power off/on the tube data must be input.
When ordering a new Zero power RAM, supply the serial number of the
generator (you will find it at the back) so the RAM can be programmed in
the factory.
The Zero power RAM is normally installed by a service engineer. If you
would like it to be installed by your own qualified technician, this is
possible, but there is then no warranty:
-
Disconnect the mains power connector.
-
Move the generator forward by means of the spindle at the rear. This spindle
must fall in the second hole in the runner.
-
Remove the small upper cover at the front.
-
Remove the first board (in front).
-
Replace the zero power RAM.
-
Run the 590 modify option:
-
For input of time: 1 0+enter
-
For input of date: 1 1+enter
Water flow switch
The water flow can be checked with a flow meter or with a bucket with
litre lines. It should switch off below 3.5 l/min.
When the switch does not work properly:
-
The switch is found behind the draining handle: remove the plate with four
screws.
-

-
Check if the voltage on the wires is switching.
-
Remove the dismountable part (first switch off the water) and check for
pollution.
-
When there is no pollution, a new switch must be ordered (only the removable
part). Readjusting the switch normally does not help.
A new water flow switch is delivered pre-adjusted at the factory, indicated
by the lacquered switch. For an old version square flow switch an conversion
kit is available.
Internal power and fuses
-
Power indicating lamps and fuses on the power unit. Slide the generator
out of the cabinet and open the small top-front cover:
| Power LED's |
Fuses (in sequence up/down) |
| on |
+5VP |
F5 |
0.5 A slow blow |
+5VP |
| on |
+5V |
F1 |
1A slow blow |
+5V |
| on |
+12V |
F2 |
1A slow blow |
+12V, -12V, -5V |
| on |
-12V |
F3 |
1A slow blow |
+12V, -12V, -5V |
| on |
+8V |
F4 |
2A slow blow |
+8V |
| on |
+15V |
F6 |
4A slow blow |
+15V, +28V |
| on |
+28V |
|
|
|
-
Backside of the generator:
-
F1, F2: 4A slow blow for mains power of electronic
-
Automatic fuse: for mains of HV power.
Troubleshooting
Error messages
When an error occurs:
-
The HV is switched off.
-
Depending on the type of error the generator is retrying, showing the message
SYSCHECK (without the type of error).
-
After retrying the error message is shown. This message disappears after
1 minute, but the green System OK remains flashing and the error message
can be can retrieved by pressing enter.
-
If the error condition is removed, the message must be cleared by pressing
clear, before operating can resume.
-
As soon as an error occurs, it is saved in the error history. Because an
error can be caused by preceding errors (and only the last is shown) it
is advised to always read out the error history
A selection of the error messages. For other messages contact the Service
department.
MA MAX
X-ray tube defect (high voltage shortage between anode and cathode).
UDC MIN
-
Automatic "Stotz" fuse at the back of the generator in the OFF position.
-
Mains power is too low (below 180V) or unstable (when immediately after
starting X-ray).
-
Check in the error history if this error is preceded
by another error (the real error, which switched off UDC)
-
X-ray tube is defect.
-
Please check and inform the service department:
-
Measure the UDC in 590 mode (27). It starts at zero.
-
Switch on HV and describe exactly what happens:
-
Do you hear a large relay switching on
-
Is UDC rising or not (should be about 350)
-
Is the automatic fuse switching off? If it is, does it switch off immediate
after switching HV on or after a short time.
... V MIN
Internal power voltage is too low. Check the power
supply
UFIL MIN
Filament voltage is too low. Check the +15V/+28V supply.
CAB TEMP
Temperature inside the cabinet is too high. Check the fan in the Power
unit (with the red lamps).
OIL TEMP
Oil temperature inside the HV tank is too high.
-
Cooling water temperature too high.
-
Cooling water flow too low.
OIL LEVEL
Oil level inside the tank is too low. Check for oil leakage in the cabinet.
CABLE LINK
The dummy connector at the HV cable is not placed in the connector on
the X-ray tube, indicating that the HV connector is not safely connected.
SHUTTER
The shutter is not fully closed (according the shutter switch the shutter
is open, while is should be closed).
EXT XLAMP
The external X-ray on lamp on the X-ray tube is defect.
-
Check if the GON UNIT cable between FR590 and goniometer is connected
(the X-ray lamp is electronic driven by the Goniometer Interface.
-
Check the power inside the Goniometer interface (indication LED's on the
back plane when the Interface is opened)
WATER
The cooling water flow is below 3.5 l/min.
RAM ERR
Turn the generator off and on again. If no error is detected this time,
normal operation is possible after confirmation of the previous RAM error.
SPARK
There were more then 255 HV flash-overs within an hour (it was about
6 in the past and in older manuals).
-
Clean the HV connector in the X-ray tube.
-
X-ray tube is defect.
COMP ERR
The real kV and mA differ from the setpoints chosen.
-
No mA. After X-ray on there is kV but no mA (shortly thereafter
the kV slowly decrease). This indicates a bad connection with the tube
or an absent filament current.
-
Check if not a short tube was mounted with the long X-ray tube spacer.
-
Clean the contact of the HV connector to the tube.
-
Be sure the brass connector set is pushed in the connected inside the rubber
tube. Remove and mount again
-

-
Pull out the brass connector and check if the connections inside are all
right, also when pressed. If not, a new part should be ordered.
-
Measure the resistance between the two connections of the tube (middle
and ring, it needs long test probes). If should be low ohms. If not, the
tube (filament) is defect.
-
When a tube was overheated
-
Defect X-ray tube.
-
A comp error can also be caused by preceding other errors, so always read
the error history.
-
Please check and inform the service department:
-
Measure the UDC in 590 mode (27). It starts at zero.
-
Switch on HV and describe exactly what happens:
-
Do you hear a large relay switching on
-
Is UDC rising to about 350
-
Are the setpoints rising to 20 kV, 5 mA?
-
Are the outputs rising to 20 kV, 5 mA?
DRAIN ERR
The drain valve is open during HV generation.
UNCALIBRATED
The battery inside the zero power RAM (IC 13 on the Control board) is
empty. Replace the ZRAM
It is possible to continue working with the generator by entering,
without change, all unknown parameters (do input the maximum voltage, current
and power), but this must be done every time the generator is powered up.
Tube broken with water
leaking
An old tube in which a hole is burnt in the anode. The leaking cooling
water will have broken the hot glass.
Overheated tube
When for some reasons the tube has been overheated:
-
The HV connector should be inspected for damage by heat. If the damage
is not visible worse, check like described in Comp
errors. If it is worse, new parts could be needed.
-
Check all other parts inside the tube holder for damage.
All kinds of error messages
With all kinds of not related error messages the cause can be electrical
disturbance from HV flash over or sparks.
-
Try a new tube.
-
Check the HV cable on spark problems
-
A combination with GX13 is always more susceptible for electrical disturbance.
-
Shutters can be opened and closed with the clock or manual buttons. The
LED's reflect the active state. A shutter can only be opened when the following
items are fitted in place and are working properly:
-
Collimator
-
Radiation safety enclosure doors closed. Alternatively the doors can be
open but a flashing warning device must be used.
-
Shutter lamps
-
Pause mode
-
Led blinking when "HV ON" is started.
-
30 seconds delay.
-
All shutters are closed.
-
Clock time is at halt.
-
The remote shutter control box can be used.
-
Active shutters re-open en countdown is resumed when the pause mode is
left
-
Manual mode
-
To enter: leave the clock/set point or pause mode first.
-
When active the shutter will be opened and the timer starts counting up
until the maximum of 199.59.59, then the shutter closes.
-
When there is a shutter problem the shutter remains closed and the indication
is "OFF"
-
Clock mode:
-
"ON" during clock mode.
-
"OF" when time has elapsed.
-
When activating the clock mode the shutter will be opened and the timer
starts counting down until 000.00.00, then the shutter closes.
-
When there is a shutter problem the shutter remains closed and the indication
is "OFF".
-
An error message is cleared when you leave the clock mode.
-
Deactivating the clock mode will close the shutter end restore the clock
set point
-
Set point mode
-
"ON" during set clock mode and set point recall mode.
-
The set point button is used to recall the set exposure time
-
Set clock
-
Only active for one shutter at a time. Two functions are possible:
-
Initialize timer set-point (clock mode deactivated)
-
Change time during experiment (clock mode active)
-
Display: Maximum time 199.59.59 (hhh.mm.ss) for clock and manual
mode. Timer input is at the blinking digit. Invalid input is ignored. Timer
fields can be left by pressing a dot. You exit the set clock mode by pressing
<enter>
-
Led bar: Extinguishes gradually during the last 60 minutes.
Testing the displays and LED's:
-
Press the key sequence: 5 9 0 enter
-
The display will show: enraf nonius cu 590 rel 1.0
-
Press key 1: all displays of each subsequent display section are on
-
Press key 2: display driver test: the number 8 walks across the display
-
Press key 3: segment driver test: the segments rotate
-
Press key 4: led bar test
Errors
-
ERROR1
-
This is an error of minor importance and is kept local to the FR561 tube
shield controller.
-
No more attempts are made to open the shutter.
-
The error message can only be cleared by remedying the error and subsequent
repetition of the action which produced the error message.
-
When a attempt has been made to open a shutter and the safety feedback
is negative then the ERROR1 message will appear in the display. It can
be caused by one of the following reasons:
-
Defective shutter lamp (even one)
-
Defective shutter: the shutter sticks in closed position.
-
Radiation safety enclosure jumper connector not installed.
-
Radiation safety enclosure doors not closed.
-
Radiation enclosure flashing warning device defective.
-
Collimator not installed.
-
ERROR2
-
The ERROR2 results from an unsuccessful attempt to close a shutter and
shutter is not fully closed.
-
The FR590 generator will shut down the HV and put the FR561 tube shield
controller into a SYSTEM PAUSE for as long as the error conditions remains
active.
-
The error message remains on the display until the FR561 tube shield controller
detects that the defective shutter has been closed.
Technical data
Manufacturer
and Service
BRUKER AXS B.V., P.O. Box 811, 2600 AV Delft, The Netherlands, Tel:
31 15 2152400 Fax: 31 15 2152500.
Service: see our web site: www.bruker-axs.com