FR591
June  2010
    Technical user manual


Introduction

The FR 591 rotating anode X-ray generator consists of a high voltage supply and an X-ray tube with a rotating anode. The hot filament in the cathode cup emits electrons that are accelerated by the high voltage. The bias voltage together with the shape of the focus cup focuses the electron beam in a small focal spot on the anode. The impact of the electrons on the anode generates X-rays of an energy/wave length depending on the type of material of the anode (Cu : 8 keV / 1.55 Å, Mo : 17,44 keV / 0.71 Å). Only a very small portion of the energy of the electrons is converted to X-rays, the rest of the energy is converted to heat. The rotation of the anode spreads the heat over a large area. This makes it possible to use higher loads than on tubes with stationary anodes. The anode rotates in vacuum and is internally cooled with water. A magnetic fluid seal around the anode shaft maintains the vacuum while rotating. Continuous pumping by a turbo molecular pump backed by a pre-vacuum pump maintains the high vacuum.

The generator consists of :



Safety instructions

Alterations, other than mentioned in this manual, can obstruct the safety precautions and are therefore not recommended. This includes adjustments which are sealed. Parts may only be changed by identical parts.

X-ray safety

The equipment can generate open X-rays, which can be hazardous when safety rules are not followed:

Beryllium

The X-ray tube contains windows of beryllium. In normal use these area's cannot be reached, but during maintenance, care must be taken not to touch the window. When by accident the beryllium is touched, wash with soap and water.

Fumes and dust from beryllium metal can be hazardous if inhaled. During use, corrosion of the beryllium window may occur, but it should not be cut, machined or otherwise removed. Disposal must comply with the national and local regulations governing beryllium.

Electrical safety

Access to the area under the table is only allowed by qualified technical personnel. Power terminals can be reached there on the outside of all modules. When a qualified technician wants to access these parts, first disconnect all automatic fuses and disconnect the mains supply before accessing any interior part of the generator.
 


Specifications
X-ray
Voltage 5 - 60 kV
Current 9 kW version : 2 -150 mA 
18 kW version : 2 - 300 mA
Stability Voltage and current : 0.01%
Power (kW) 9 kW (1 modules) or 18 kW (2 modules) 
Focus Type Filament W Cu Mo Ag/Au Fe/Cr
1.0 x 10 mm BF3 broad 15.0 15.0 15.0 15.0 15.0
0.5 x 10 mm BF2 broad 15.0 15.0 15.0 11.0 9.0
0.3 x 3 mm FF3 fine 8.2 5.0 5.0 3.1 3.0
0.3 x 3 mm HBA FF3 fine 6.0
0.2 x 2 mm FF2 fine 4.5 3.0 3.0 1.7 1.8
0.1 x 1 mm FF1 fine 1.6 1.2 1.2 0.6 0.6
0.1 x 1 mm HBA FF1 fine 1.5

0.5 x 5 and 0,1 x 2 are in the list but are not configured, so cannot be used 
HBA = High Brilliance Anode

Focus Focus types : see Power
Stability focal spot position <10 microns peak to peak
Focal height 522 m above working space, optional 300 mm
Window 0.25 mm beryllium
Take off angle 0° - 10°

The shutter does not open, or is closed when: The X-ray high voltage is not switched on, or is switched off when: Safety override
The system is supplied with two overrides. Their use should be limited to technical maintenance (see safety rules). Safety signals With all lamps, a defective lamp is detected and activates the safety circuit is activated.

Environmental, according to IEC 1010
Environment Indoor use
Altitude Up to 2000 m
Temperature, operating °C (within specification)
Relative humidity Maximum 80% if <31°C, decrease linearly to 50% at 40°C
Pollution degree 2, according to IEC 664 (normal clean environment).
Protection IP 20, according to IEC 529 (limited protection to water).
Gasses No explosive gasses are allowed.
Heat generation to air 9 kW : max. 1 kW; 18 kW : max. 2 kW
HV tank oil Shell Diala D. The oil does not contain  PCB's. This oil is also used in FR571, GX21 and FR58.. generators
Power supply Nominal 200V-240V (phase - neutral), 347V-415V (phase =- phase).
Fluctuations ±10%
Frequency 50 - 60 Hz. ±5%
Connection Three phase Y (WYE or star), five wire (3 line, 1 neutral and 1 ground). No earth leakage relay / protectors can be used.
Mains fuses 9 kW : 3 x 35 A; 18 kW : 3 x 63 A
Power (kVA)
L1 L2 L3 Total
9 kW version 4.3 4.7 6.2 15
18 kW version 7.8 8.2 9.7 26
Installation category II, according to IEC 664.
Water cooling
Flow and temperature
  • Normal Cu or Mo anode (until 1998):
    • min. 15 l/min (9 kW, input 15 ºC) or min. 25 l/min (18 kW, input 15 ºC).
  • New Cu High Power Anode:
  •  
  • Minimum temperature : +15 ºC ( to prevent condensation)
Water
  • Primary water : normal tap water quality.
  • Secondary water : demineralized water.
Input pressure 2 - 6 Bar
Nitrogen Advised for a clean vacuum system. Max. 10 l /year. Pressure: 0 - 0,3 bar
Weight 967 kg


Operation

Start up procedure



Shut off procedure



Shut down

There are several ways to shut down the generator :



kV/mA set



Control panel.  purple: only in Microstar

Operate Setup Diagnose Service View Status
          .
.
bits line
.
 .
StdBy HotKey shutter 1 shutter 3 X-ray FR591, MICROstar

Buttons/switches:

Operate
Generator
  • Voltage, goal: set kV
  • Current, goal: set mA
  • Power, goal: shows kV * mA (no setting)
  • Bias, goal: change the precalculated or set  (hatched) default value. After HV off/on it will go back to the default value.
Recrystallize Starts the automatic recrystallization program (ignore "done" before start). It runs an increasing current through the filament without HV during about 30 minutes. This is done after mounting a new filament. Recrystallize will only start when the vacuum is below 1E-6 mbar.
Conditioning Starts an automatic power up to clean the tube from gases by heating up. It rises slowly the power till the kV/mA limits, while checking the vacuum and trips. If vacuum is too bad or too much trips it slows down the powering up of even goes down. This is done after mounting a new filament after recrystallize. It can take several hours. If the tube and filament are clean, it can be done in about half an hour.
Anode Speed To choose a higher then default calculated anode speed. This is used for checking an anode on high speed without HV. The real speed will be 10% higher then this setting.

 
Setup Setup parameters
Shutters Who is opening the shutters:
  • GCU enabled (Hardware) : goniometer can open the shutter through a (GCU) cable with hardware lines. It is not possible to open or close the shutter from the control panel 
  • GCU disabled (Software) : the Control panel can open the shutter. 
Generator Power modules used: number 1, number 2, or 2 modules (18 kW). Don't change this. Microstar: fixed on module 1
Tube Be sure you have the management over the X-ray (Setup • Management) and the generator is in status Power up
  • Focus size: Microstar: fixed on 0.1x1
  • Filament: fine (default) or broad. Microstar: fixed on fine
  • Target: Anode material. Microstar: fixed on Cu. This setting can also be used to reset the safety circuit of the tube covers: choose the anode which you use and enter. This will reset.
  • Limits: user limits of kV, mA and power. The absolute maximum (tube) limits are calculated from the focus, anode, power modules etc.
Management
  • X-ray: permission to set HV.
  • Service: permission to write data in the electronics. Only for service. Default: No management
  • Shutter: permission to open the shutter
  • Choices:
    • Obtain: Control panel takes the management.
    • Release: Control panel gives the management away to the external RS232 communication line.
  • Display:
    • Panel: manual control by the Control panel
    • No management: Management is released from the Control panel and ready to be taken over. When this happens it shows:
    • RS232: management by software trough the RS232 connection. This can be the Server-, BIS-  or service program.
Preferences
  • Date/time: Type of date (European, American). 
  • Units: British (mbar and °C, default), USA (psi and °F) or ISO (torr and °C). 
  • Contrast (of display) :18 - 22 typical. Be careful! When choosing a different value the display can disappear and this adjustment must be found in the blind.
  • LeftbuttonShutter: the left button drives shutter 1 (default) or 3.
  • RightbuttonShutter: the right button drives shutter 1 or 3 (default).
  • Preferent bias: the default bias value
(RS232) RS232 settings cannot be changed in Setup. Better change in Service

 
Diagnose only in Microstar
Error logs History of warnings, faults and errors. It is however easier to use the service program for this
Trips Number of trips since power on (not permanent reading).
Tube Switches To reset the cover safety circuit. When tagged it should show 7, indicating that all switches in position save. If not, then the program will ask if you want to reset the safety circuit. After agree it will be reset and "7" will show (unless one of the switches is still not in save position).
Observe Shows some internal values. Only for service engineers. 0 = filament current, 1 = frequency, 2 = Pib

 
Service Special service commands, only when asked for by the Service department. For writing (in Microstar: those who are hatched): first get management.
Read Read of memory on request of the Service department, who also will give the settings (nowadays mostly the service program will be used):
  • Node: type of board:
    • Main-io: Main board, IO-processor
    • Main-ctl: Main board
    • Tube-ctl: Tube board
    • Gen-ctl: Generator board
  • Area: RAM (default) or code (EPROM)
  • Address: $## ##
  • Bytes: 4 (default), Permanent (default) or single
  • Format: Hexadecimal (default), decimal or binary
  • $####  - $##: shows the 8 bits data on the address requested and three following addresses
Modify Write to the RAM memory. Be careful: wrong writing can put the system in a dangerous status, resulting in damage. 
IO Reading and writing of IO registers. It is a combination of Read and Write but only for registers, so with address $ff ##. Be careful : wrong writing can put the system in a dangerous status, resulting in damage. 
Show revisions
  • Hardware: boards version. This is set on the board with resistors.
  • Software: firmware version in EPROM
Ventilate To enter nitrogen or filtered air in the tube, so it can be opened. Opens the nitrogen NC valve for approx. 30 seconds. This must be done several times before the cathode can be removed.
what to toggle to set or reset four User (main) relays. These relays can be used by customers to switch their equipment on/off. In practice they are seldom used.

 
View Show parameters (not adjust)
Primary status Shows the actual kV, mA, power, bias,  anode speed and vacuum
Setup parameters Shows as mostly set in Setup and can be better checked there.
  • RS232 parameters are not right. You can better read in Service
  • Shutter: 
    • "GCU...": see Setup
    • "key switch": the shutter key switch is off.
  • 9 kW modules used
  • Anode, focus, filament type.
Operation parameters
  • on/off of : water pump, oil pump, anode motor and vacuum pump. Status:
    • Stopped: off
    • Started: has been switched on. Waiting till it is ready (flow presnt, vacuum OK, etc)
    • Unknown: 
    • Polluted (vacuum): 
    • Ready: running alright, which a check on flow, speed, vacuum. All should be in this status before the generator is in full Standby
  • value of: temperature of oil, primary input watersecondary output water and anode output water.

 
Status Shows the (intermediate) status or the intermediate status
  POWER UP : only power on 
STANDBY : all equipment running and ready to generate X-ray 
GENERATE : generates X-ray, HV is on.
#1 - #2
    #
To  #1 - #2
going from status #1 to #2. Showing time bar
stable at status #
First fallback to #1, then build-up to #2

sec/10   sec 
              sec  sec/10
Time line of transition. sec = seconds, sec/10 = in 10 sec units
command delay time
condition time out

 
Shutter To open or close shutter 1 and 3
Buttons
  • If the shutter is setup to be opened by the goniometer it is not possible to open from the control panel, but because of safety it is always possible to close the shutter with these buttons.
  • The opening of shutters can be disabled by safety key switches under the X-ray tube, to prevent unauthorized opening of the shutter.
Display
  • FR591
    • "Open",   "Close": function of the button (not the status of the shutter). So "open" means that the shutter is closed and can be opened by the button.
  • Microstar:
    • "Open 1, 3",   "Close 1, 3":  Panel is manager. Function of the switch: "Open" means that the shutter is closed and can be opened by the button.
    • "1, 3  Open",   "1, 3  Clos'd":  External is manager. Status of the shutter.
    • Errors: see Shutter error

 


Maintenance



General warning

The mechanical maintenance described is intended for customers who have attended a maintenance course. It comprises only a part of the required periodic maintenance. The skill of an experienced service man is required for items related to high power, delicate mechanics and vacuum. When problems arise with the described items please first ask Service department  for advise before using brute force. Bruker AXS cannot give free technical support for work outside what is described in this manual and is not responsible for the work, done by the customers.
 


Periodic maintenance

For a long lifetime of the generator a yearly preventive maintenance is strongly advised. Some moving parts have a limited life and can damage expensive parts when not replaced in time. A maintenance by Service consists of:

Some items, mentioned in this manual, can be done by the customer after a maintenance course, but there are limitations:



Filament replacement

Filaments are supplied in a set together with mounting rod screws, clamp segments, Allen keys and cathode O rings.
 

General
Preparations
Fine focus
Broad focus
Cathode mounting
Recrystallization


There is no tool for a more optimal horizontal alignment of the focus. The previously used pin hole with fluorescent screen is not allowed anymore because of the X-ray background radiation. When you want to use your own safe tools, for instance using film inside a radiation enclosure, you should get permission from the local radiation safety officer. The radiation enclosure is designed to suppress only the weak scattered radiation and not the high intensity of the primary beam.

When there is a possibility to view the focal spot, the shape will be:


 


Water seal replacement

The water seal in the water bridge needs to be replaced at least every year.

New High Power Anode



Shutter replacement



Anode drive belt replacement

The anode drive belt should be replaced every year.



Anode replacement for Normal Power Anode (NPA)
 

General
Rotorplate remounting
Anode disassembly
Anode cleaning
Anode remounting
Rotorplate mounting



Anode replacement for High Power Anode (HPA)
 

General
Rotorplate remounting
Anode disassembly
Anode cleaning
Anode remounting
Rotorplate mounting
Anode speed sensor



Ferro seal replacement for Normal- or High Power Anode



Recoil shield cleaning



Anode motor brushes

The four brushes should be replaced every year or when they are less than eight millimetres long. (when new they are 23 mm long)



Vacuum measuring probe for FR591 (red type)

The clamp under the vacuum measuring probe can be opened to enter air into the head.

The vacuum measuring probe can be inspected when it does not start up. Do this only when asked by the Service department. One can remove the measuring probe by rotating it anti clockwise and lift (bayonet) without loosening the clamp.

Checks to get the vacuum measuring head starting:

When remounting, check if the probe is centred properly.


Pre-vacuum pump oil change for FR591 pre-vacuum-oil pump




 

Shutter replacement or exchange with shutter 3



Safety test

Shutter safety, with the FR590 or FR591.

  1. Switch off the X-ray high voltage
  2. Open the shutter. Check both shutter-open lamps.
  3. Check if the shutter will not open after activating the next unsafe conditions :
  4. Check the flashing Open-beam lamp when the radiation enclosure safety (Open beam) is override by the key-switch.
X-ray high voltage safety
  1. Switch on the high voltage to minimum: 20 kV 5 mA
  2. Check the X-ray on lamps.
  3. Check if the high voltage falls off when:



Technical data

Cooling, vacuum FR591

Cooling, vacuum Microstar


 
 

Microstar (Ultra)

FR591

 
Position of (anode) water flow switch

in Microstar, FR591 new , FR591 old

FR591

Electrical overview


Part number list

Next items are advised to keep on stock or buy before yearly maintenance. The table gives the numbers delivered with a new system in the spare parts kit and the yearly use. Check if the part is for the normal power anode (NPA) or the new high power anode (HPA). For expensive components it is advised to buy when needed, to prevent the part is out of warranty when it is used.

The next items we advise to keep on stock:
 
Kit Use
 yr
Part number Description
5 0571842
0571843
Fine focus filament set (5x) including mounting parts and tweezers or
Broad focus filament set (5x)
2
1
1 2985013
2134044
Water seal in water bridge, the standard replaced item for NPA or:
Water seal for HPA
1 1 3590003 Anode motor brushes, set of 4
2 1 2692023 Brush for anode grounding in the waterbridge
5 1 4010071 Silicone oil and  washer for the HV connectors
5 2603050 Shutter lamp, inside the shutter. 5 V 60 mA
10 5 2603335 Shutter lamp, on the head. 24V 50 mA
10 5 2603212 X-ray on lamp. 12V 120 mA
5 2603235 Open beam lamp. 12V
1 4000014 High vacuum grease for O-rings
0591114R Shutter, revised. Optional. This is an expensive part, but take into account that because of the high radiation environment and with high use the shutter is vulnerable to wear and for safety reasons it is not possible to use a generator with one shutter missing, also on a port which is normally not used.

For customers who execute the yearly maintenance (partly) themselves, we advise the next items will be bought:
 
Kit Use
 yr
Part number Description
1 Depends
on type
Anode. It is possible to
  • Send the anode to the factory for refurbish (grind when needed, polish, internal chemical cleaning)
  • Buy an anode and return the old anode.
  • Buy a refurbished anode and return the old anode. This is only possible when on stock.
The lifetime of an anode is five years.
1 0571355 Ferro seal.
1 1 0571319
0168374
Counter face water seal, on the anode axis for NPA or:
Counter face seal for HPA
1 1 2033114 V-belt
1 1 4010021 Oil for the pre-vacuum pump, 1 ltr
5 2 2132162 O-ring 17.10 x 1.60 mm, for the anode pulley and shaft
5 1 2132163 O-ring 19.10 x 1.60 mm, for the pulley face seal
5 2132310 O-ring 31.42 x 2.62 mm, for the water bridge for HPA
5 2132429 O-ring 37.77 x 2.62 mm, for the water bridge for HPA
2 2132528 O-ring 56.82 x 2.62 mm, between measuring flange and head
5 2132431 O-ring 47.0 x 5.34 mm, for cathode 
2 1 2132529 O-ring 58.42 2.62, for the recoil shield
2 1 2132634 O-ring 66.35 2.62, for the recoil shield front
5 1 2132963 O-ring 94.97x1.78, between the rotor plate and head
1 2140073 Grease cartridge for the turbo pump


Technical warnings
 


Disposal
 


Troubleshooting

Startup errors Low vacuum Shutter does not open. (sometimes error 13-08) Errors mentioned on the control panel

Selection of errors from the control panel. For other error message contact the Service.

Look at the second numbers of the error, for instance (0x13, 0x08) is 13-08 :
12-06
12-40
12-41
12-43
12-44
13-01
13-02
13-03
13-07
13-08
13-0B
13-0C
13-12
13-13
13-16
13-43
13-56
14-0A
14-0D
14-0E
14-42
14-4B
14-4C
14-55
14-56
Errors without error number: 

12-06 EMERGENCY PUSHBUTTON

The red emergency button on the outside of the cabinet is pressed

12-40 NOT STANDBY FOR HV

Not ready to generate HV (Generator is not in Standby status).

12-41 kV NOT RECHARGED

Action taken, for instance HV ON, when the kV wasn't fully discharged after the HV OFF. Wait some seconds and repeat.

12-43 BAD VACUUM RECRYST

Starting recrystallization when the vacuum is too low. Wait till the vacuum is high and the recrystallization will continue automatically.

12-44 AUTOMATIC FUSE AUX (Microstar)

The automatic fuse of the auxiliary unit (with pumps etc.) is off

13-01 LOOSE HT CABLE END

The "HT cable present" connector on the tube is not placed.

13-02 LOOSE TUBE COVERS

The white tube cover with the red lights is not in place, detected by a reed switch with magnet in the cover. The signal is latched in a battery powered circuit, so even if the covers are in place again or power has been off, the electronics will show this error. The only way to reset is to choose the (same) anode target (Cu or Mo) again from the control panel.

13-03 LOOSE TUBE SHIELD

The connector of the tube shield is not placed, indicating the metal tube around the cathode of the X-ray tube is not in place. The signal is latched in a battery powered circuit, so even if the connector is placed again or power has been off, the electronics will show this error. The only way to reset is to choose the (same) anode target (Cu or Mo) again from the control panel.

13-07 TEMP ALARM COOLWATER

The temperature of the cooling water at the secondary output of the heat exchanger is too high. Check the input water flow and the temperature. The temperature is shown in menu View / Operating parameters.

After restoring the cooling the temperature of the secondary water can still be too high, preventing the water pump to start. It has to cool down on its own, which can take some time. It is  possible to start the water pump by a service command. Use this command only when needed, because wrong use can start other parts without safety.

13-08 SHUTTER FAIL

One of the shutters doesn't open, doesn't close or the safety shutter fitted switch is not activated.


13-0B ANODE DRIVE OVERRUN (FR591) or ANODE DRIVE CURRENT (Microstar)

Start again and go in smaller steps to the kV/mA power.

13-0C VACUUM GAUGE FAILS

Vacuum gauge doesn't work. Possible causes:

13-12 LEVEL ALARM WATER

Cooling water level in the vessel is too low. Refill the vessel with demineralized water (first read electrical safety)

13-13 FLOW ALARM TUBE

No water flow through the anode. Causes:

13-16 VACUUM ALARM

Check the low vacuum list

13-43, 56  TURBO SPEED FAILS or TURBO PUMP TIMED OUT

Turbo pump speed too low or Turbo pump did not reach the right speed after 4 minutes.

14-03 TEMP ALARM HEATSINK1

The heatsink is located in the power module, situated left of the HV tank. This is cooled by two thin brown tubes.

14-0A LEVEL ALARM HV TANK

The oil level in the HV tank is too low. Check if there is oil leakage on the floor.

14-0D XRAYS DISABLED

Trying to generate X-ray while the X-ray key switch is in position disabled.

14-0E XRAYS LAMP FAILS

Both red X-ray On lamps are defect.

14-42 FILAMENT SET FAILS

This can be caused by:

14-4B GATE DRIVE FAULT 14-4C FILAMENT BROKEN (FR591) or FILAMENT CHECK FAILS (Microstar)

Filament current too low, most probably a broken filament

14-55 AUTO

The Stotz switch of a 9 kW module is switched off. Switch it on again.

14-56 SHORT

A short circuit between the filament and the bias cup. The filament has to be replaced. If it is caused by a rotation of the filament, then check with the next filament that it does not change during fastening (see description  filament replacement)

Internal error: Input buffer overrun

This error is a communication error on the digital bus and is mostly caused by electrical disturbance. This can be:


Communication problem with computer or goniometer

Anode motor brushes wear

The anode brushes should last at least a year. When they wear in a too short time you could clean the motor inside and  track the length for some time. When new, the brushes are 23 mm long. If cleaning does not help, probably the motor is worn and should be replaced.

No mA

Open the tube and check if the filament does not touch the focus cup.

Service program

A service program is available which reads the internal registers and values for trouble shooting the FR591. The program runs on a PC under Windows95/98.



Manufacturer and Service

BRUKER AXS B.V., P.O. Box 811, 2600 AV Delft, The Netherlands, Tel: 31 15 2152400 Fax: 31 15 2152500.

Service: see our web site: www.bruker-axs.com