The standard way to operate the generator is through the control panel.
Note that: the display shows the function of the buttons (the next action),
not the present status. So open in the window means that the shutter will
open when the button is pressed.
| Setup |
|
| Shutters |
Shutter (1 or 3) control:
- Software: goniometer can use the shutter. It is not possible to open
or close the shutter from the control panel
- Hardware: Close the shutter from the control panel. With special
electronic strapping also : open shutter. |
| Generator |
Fixed at module 1 |
| Tube |
Fixed at 0.1x1 focus, fine focus, anode Cu, power 2.7 kW
Be sure that you have the management over the X-ray (Setup • Management)
and the generator is in status Power up. |
| Management |
X-ray and Shutters default : Panel . When an external program wants
the X-ray management it takes over automatically and : "No management"
or "RS232" will show.
Service default : No management. Only when asked by Service department. |
| Preferences |
- Type of date: European, American
- Type of unit : British (mbar and °C, default), American (psi
and °F), ISO (torr and °C).
- Contrast (of display) :18 - 22, typical. Be careful! when choosing
a different value the display can disappear and this adjustment must be
found in the blind.
- Left button shutter: button drives shutter 1 or 3
- Right button shutter: button drives shutter 1 or 3 |
| Service |
For striped items: first get management. |
| Read |
Read of memory. It will default give the data of four successive addresses
in hexadecimal format, starting with input address (which will be supplied
by Service department) |
| Modify |
Write to the memory. Be careful : wrong writing can put the system
in a dangerous status, resulting in damage. |
| IO |
Reading and writing of IO registers. Be careful : wrong writing can
put the system in a dangerous status, resulting in damage. |
| Show revisions |
Show board revisions, adapted at the board with resistors |
| Ventilate |
To enter nitrogen in the tube. Opens the nitrogen NC valve for approx.
30 seconds. This must be done several times before the cathode can be removed |
| What to toggle |
To set/reset User (main) relays |