Kappa(X8)APEX logo
November  2008

 
    Technical support


Safety instructions

X-ray safety

Safety rules as addition to the User Manual:

Beryllium

The detector and X-ray tube contains windows of beryllium. In normal use these area's cannot be reached, but during maintenance, care must be taken not to touch these windows. When by accident the beryllium is touched, wash with soap and water.

Fumes and dust from beryllium metal can be hazardous if inhaled. During use, corrosion of the beryllium window may occur, but it should not be cut, machined or otherwise removed. Tube and detector disposal must comply with the national and local regulations governing beryllium.
 


Maintenance

Regular maintenance and checks
 

Cleaning

For cleaning use a damp cloth, moistened with water or isopropyl alcohol. No cleaning agent may enter the goniometer.


Tools


 

Server program

If the server program window is not visible it can be restored by editing a new line with SHOWWINDOW in C:\program files\brukeraxs\bcp and pis\KCCD.SET
 

Safety test

This test it limited to some safeties in the goniometer without an internal failure check:


X-ray optics alignment

The measurement is done in a less safe condition: without beamstop and the enclosure safety override. Safe operation is possible when executed by a qualified person and precautions have been taken (see safety instructions)
  1. Don't let other persons come in the neighbourhood of the system.
  2. Don't leave the system unattended, before reestablishment of the safety.
  3. Don't come in the primary beam, although for security reasons it is attenuated to a weak and safe intensity.
Measurement set-up
  1. Service key on and door lock connector on the service panel.
  2. The main shutter can be opened by pressing the red shutter button on the service panel for 5 seconds.
  3. Place a #353 collimator
  4. Place the absorber on the collimator (take care: Mo and Cu use different absorbers).
  5. Remove the beamstop and place the red vane (safety switch override).
  6. X-ray generator:
  7. For measurements use in BCP: Tools, Halfbeam
Adjustment
  1. Adjust Tube 2Theta to 0°.
  2. Dismount the monochromator and mount the X-ray port (two Allen screws {2.5 mm} at the bottom, which can be reached by two holes {not the middle one} under the black monochromator holder). During replacement the power of the X-ray generator must be off.
  3. Adjust tube 2Theta = 0° and take-off = 3.0° (for fine focus 0.4 x 8 mm).
  4. Adjust on approx. maximum intensity with tube vertical and tube horizontal adjustment (measurement Continuous)..
  5. Mount the half-beam absorber (3 mm bar) on a goniometer head. Align it with the microscope with and Phi 90° rotated: the edge of the bar must be in the centre (it must absorb half of the beam).
  6. Align horizontal and vertical:
  7. Test the alignment of the half-beam absorber :
  8. Remove the half-beam absorber and X-ray port. Switch off the power of the X-ray generator during replacement.
  9. Mount the monochromator and switch on the power of the X-ray generator.
  10. Adjust Tube 2Theta (Mo: 12.2°, Cu: 26.6°) and the Monochromator Theta on the anode target material used.
  11. Adjust the intensity on maximum. First with monochromator 2Theta, then tube 2Theta and monochromator height.
  12. Lock the monochromator height.
  13. Measure the open intensity.
  14. Mount the half-beam absorber and position horizontal with "Vertical" (don't use the result of this measurement). Measure continuously and adjust with monochromator theta till the intensity is half the previous measured open intensity ±10% (a first coarse alignment).
  15. As before, adjust the intensity within 5%:
  16. Lock the monochromator Theta.
  17. Set the X-ray generator to 10 mA and measure the open intensity (measurement Open) without the half-beam absorber. This is the Effic value.
Adjustment points


EFFIC: intensity of the primary beam

The Effic measurement is a relative measurement of the intensity of the primary beam, which measures the total efficiency of the X-ray: tube, optics and detector gain. When the value is too low it is mostly caused by the tube (degradation) or the X-ray alignment. The measurement is done during installation of the system and later measurements can be compared with this value. Take into account that because of tolerance in collimator and absorber (material thickness) the values cannot be accurately compared with other systems.

Measurement:


Bootstart BIS remove


Troubleshooting

When consulting Service always run D8tools, make a status file (.sts) and send it to the service center


Serial number
 
A service department will often ask for the system serial number. All parts has its own serial  number, but also the whole system has a serial number. It is a number 1-9999 (from Madison factory) or a 4.... number (From Delft). It can be found on one of the three places:

C) The right side post just below the leds. (Most probably)
B) The right top of the rack underneath the enclosure.
A) The top left brace behind the glass.

Using D8tools

D8tools is the program to check for errors in the X-ray dafety en shutter circuit. Start:


Resetting HV errors


Start-up, PC and communication problems


Shutter does not open


HV goes off


Collision and maximum angle switch

The anti-collision software prevents the goniometer from colliding with the collimator holders or the axis to run past their limits. When a collision takes place because of unexpected circumstances, like a technical error, it is sometimes possible to return:

  1. On the Server computer a menu appears.
  2. Choose the axis and direction to run out of collision.
  3. Press the overrule button at the back of the goniometer. The axis now positions a small step.
  4. If not enough: repeat. As soon as the collision switch is released, the goniometer starts to relocate the zero position.

X-ray intensity or images problem


Detector problems

Make two uncalibrated images:
- Start BCP in master mode
- Choose: Tools, test frame
- Make an image: 0.1s, 1 exposure, no shutter, still, no dark&flood, file name dark0.gadd
- Make the same image with 600s, file name dark600.gadd
- Send both images (in: Program files/BrukerAXS/BCP and BIS/calibrations/) to the service centre in zipped form (without zipping errors can occur when transported somewhere with Linux)

Lytron cooler