| Kappa(X8)APEX |
 |
November 2008
Safety instructions
X-ray
safety
Safety rules as addition to the User Manual:
-
The equipment must be serviced and maintained by qualified personnel who
are trained with the hazards of and precaution against X-ray exposure.
-
The equipment contains safety circuits which should not be overridden,
other than supplied: the service key, door lock connector and beamstop
override vane. These overrides may only be used for service and maintenance.
-
When using the beamstop override vane (only for alignment) the system must
not be left unattended. Theta always at 0° and a low kV/mA setting
as described in the alignment instructions.
-
Safety circuits must be tested by service personnel or the safety officer
at least every 12 months, as described in Maintenance
-
The collimator must always be installed.
Beryllium
The detector and X-ray tube contains windows of beryllium. In normal
use these area's cannot be reached, but during maintenance, care must be
taken not to touch these windows. When by accident the beryllium is touched,
wash with soap and water.
Fumes and dust from beryllium metal can be hazardous if inhaled. During
use, corrosion of the beryllium window may occur, but it should not be
cut, machined or otherwise removed. Tube and detector disposal must comply
with the national and local regulations governing beryllium.
Maintenance
Regular maintenance
and checks
Cleaning
For cleaning use a damp cloth, moistened with water or isopropyl alcohol.
No cleaning agent may enter the goniometer.
Tools
-
Absorber: Zirconium (for Mo) or Nickel (for Cu) filter, placed on the collimator
to attenuate the intensity of the primary beam to a measurable and save
level. It is only allowed to used with low kV/mA settings. Specification:
-
Mo: 7 foils of 0.108 mm Zr, attenuation approx. 2400x
-
Cu: 9 foils of 0.02 mm Ni, attenuation approx. 1900x
-
Half-beam absorber: a 2 mm bar, placed in the crystal head.
-
Safety overrides, only allowed to be used during technical maintenance
(see safety rules).
-
Service switch which overrides the radiation enclosure safety. This state
is signalled by the flashing Service lamp.
-
Door lock connector which overrides the door lock safety.
-
Beamstop override vane. There are no extra warnings.
-
Test crystal: Ylid
Server program
If the server program window is not visible it can be restored by editing
a new line with SHOWWINDOW in C:\program files\brukeraxs\bcp and pis\KCCD.SET
Safety test
This test it limited to some safeties in the goniometer without an internal
failure check:
-
Switch off the X-ray high voltage
-
Open the shutter in Server with Acquisition,
Toggle shutter.
-
Check if the shutter not open after activating the next unsafe conditions.
The shutter must remain closed after reestablishment of the safety (except
the shutter lamps):
-
Collimator removed.
-
Beamstop removed.
X-ray optics alignment
The measurement is done in a less safe condition: without beamstop and
the enclosure safety override. Safe operation is possible when executed
by a qualified person and precautions have been taken (see safety
instructions)
-
Don't let other persons come in the neighbourhood of the system.
-
Don't leave the system unattended, before reestablishment of the safety.
-
Don't come in the primary beam, although for security reasons it is attenuated
to a weak and safe intensity.
Measurement set-up
-
Service key on and door lock connector on the service panel.
-
The main shutter can be opened by pressing the red shutter button on the
service panel for 5 seconds.
-
Place a #353 collimator
-
Place the absorber on the collimator (take care: Mo and Cu use different
absorbers).
-
Remove the beamstop and place the red vane (safety switch override).
-
X-ray generator:
-
Mo: 30 kV, 10 mA
-
Cu: 20 kV, 10 mA
-
For measurements use in BCP: Tools, Halfbeam
Adjustment
-
Adjust Tube 2Theta to 0°.
-
Dismount the monochromator and mount the X-ray port (two Allen screws {2.5
mm} at the bottom, which can be reached by two holes {not the middle one}
under the black monochromator holder). During
replacement the power of the X-ray generator must be off.
-
Adjust tube 2Theta = 0° and take-off = 3.0° (for fine focus 0.4
x 8 mm).
-
Adjust on approx. maximum intensity with tube vertical and tube horizontal
adjustment (measurement Continuous)..
-
Mount the half-beam absorber (3 mm bar) on a goniometer head. Align it
with the microscope with / and Phi 90°
rotated: the edge of the bar must be in the centre (it must absorb half
of the beam).
-
Align horizontal and vertical:
-
Start a measurement (horizontal or vertical). Two measurements are executed
with Phi 180° rotated.
-
Both measurements must be equal within 5%. If not : adjust with tube horizontal
or vertical adjustment to the proposed "new" value (with measurement Continuous)
and measure again.
-
Lock the tube vertical adjustment.
-
Test the alignment of the half-beam absorber :
-
Remove the half-beam absorber and measure Open.
-
Place the half-beam absorber and measure Vane. The error should be not
more than 25%. If it is : adjust to the advised new value and measure again.
When the alignment was bad, check the previous tube alignment again.
-
Remove the half-beam absorber and X-ray port. Switch
off the power of the X-ray generator during replacement.
-
Mount the monochromator and switch on the power of the X-ray generator.
-
Adjust Tube 2Theta (Mo: 12.2°, Cu: 26.6°) and the Monochromator
Theta on the anode target material used.
-
Adjust the intensity on maximum. First with monochromator 2Theta, then
tube 2Theta and monochromator height.
-
Lock the monochromator height.
-
Measure the open intensity.
-
Mount the half-beam absorber and position horizontal with "Vertical" (don't
use the result of this measurement). Measure continuously and adjust with
monochromator theta till the intensity is half the previous measured open
intensity ±10% (a first coarse alignment).
-
As before, adjust the intensity within 5%:
-
Vertical : with monochromator theta (with measurement vertical)
-
Horizontal : with tube horizontal (with measurement horizontal).
-
Lock the monochromator Theta.
-
Set the X-ray generator to 10 mA and measure the open intensity (measurement
Open) without the half-beam absorber. This is the Effic
value.
Adjustment points
EFFIC: intensity of the primary
beam
The Effic measurement is a relative measurement of the intensity of
the primary beam, which measures the total efficiency of the X-ray: tube,
optics and detector gain. When the value is too low it is mostly caused
by the tube (degradation) or the X-ray alignment. The measurement is done
during installation of the system and later measurements can be compared
with this value. Take into account that because of tolerance in collimator
and absorber (material thickness) the values cannot be accurately compared
with other systems.
Measurement:
-
353 collimator with Mo or Cu attenuator
-
DX = 6 cm
-
Mo: 30 kV, 10 mA
-
Cu: 20 kV, 10 mA
-
Replace the beam stop by the red vane
-
Measure the intensity in BCP: Tools, half beam: no tool, 10s (shows cps,
ADU/s, beam image)
Bootstart BIS remove
-
Check with ctrl-alt-delete. After power on, no "InstrumentServi" should
be active. If there is, remove bootstart with:
-
Start/Programs/Startup/BIS. (advise: delete) or:
-
Start-settings-controlpanel, Administrative Tools, double-click Services,
choose Bruker Instrument Service, Action-properties: automatic or disabled.
(advise: disable)
Troubleshooting
When consulting Service always run D8tools, make a status file (.sts)
and send it to the service center
Serial number
 |
A service department will often ask for the system serial number. All
parts has its own serial number, but also the whole system has a
serial number. It is a number 1-9999 (from Madison factory) or a 4....
number (From Delft). It can be found on one of the three places:
C) The right side post just below the leds. (Most probably)
B) The right top of the rack underneath the enclosure.
A) The top left brace behind the glass. |
Using D8tools
D8tools is the program to check for errors in the X-ray dafety en shutter
circuit. Start:
-
In BIS give the manual command: [manual/mode=1]
-
Start D8tools, On line status, bar-button: refresh on
-
File, Save as. Send this .sts file to the service center
Resetting HV errors
-
Start D8tools
-
Go in the left directory to: "x-ray generator / x-ray software safety circuit
capture register"
-
Select from the menu: "utilities / x-ray / reset x-ray software safety
circuit capture register" (default password = "password").
-
Wait some time.
-
Switch on the HV with the button on the side panel.
Start-up,
PC and communication problems
-
PC-server program remains in: present operation
= Goninit
-
Kappa does not move and Omega is slowly running to its maximum position:
check the dummy plug on the goniometer connector 'Limit switch'. Before
replacing first switch off/on the Interface.
-
Add at the top of the start-up file C:\Program Files\BrukerAXS\BCP and
BIS\KCCD.SET the line: SKIP_ZERO.
(take care to remove this line after testing !) After start-up:
-
Present operation = none (system has started) : It is a problem in the
goniometer. Inform the service department
-
Present operation = Goninit (system does not start) : It is a problem in
the Interface.
-
Starting BIS gives no response: Ctrl-Alt-del, Task Manager, end all InstrumentServi
processes
-
PC error messages
-
"Failed to open serial port for goniometer
controller, in use or not present", Timeout ..., "Failed to initialize
communications with the goniometer controller. This will happens when BIS
was already started before nd has occupied the comport.
-
Check with ctrl-alt-delete. After power on, no "InstrumentServi" should
be active. If there is, remove it.
Shutter does not open
-
The shutter is normally closed when
-
The collimator is not in the right position.
-
The monochromator or X-ray port is not in the right position.
-
The beamstop is not mounted.
-
The shutter open lamps are defect.
-
The radiation enclosure is not closed.
HV goes off
-
Start D8tools and check for red alarms during switching on the HV.
-
Reset errors
Collision and maximum
angle switch
The anti-collision software prevents the goniometer from colliding with
the collimator holders or the axis to run past their limits. When a collision
takes place because of unexpected circumstances, like a technical error,
it is sometimes possible to return:
-
On the Server computer a menu appears.
-
Choose the axis and direction to run out of collision.
-
Press the overrule button at the back of the goniometer. The axis now positions
a small step.
-
If not enough: repeat. As soon as the collision switch is released, the
goniometer starts to relocate the zero position.
X-ray intensity or images
problem
-
No X-ray (but no error messages):
-
Check if there is also no X-ray with the Effic measurement.
-
If the main shutter opens (audible and visible) and the led on the small
shutter is on, then probably this shutter should be cleaned
of replaced.
-
X-ray too low:
-
Check the Effic measurement and compare it with earlier
measurements. With a new tube check first if Effic without monochromator
is not too low (not possible in K3). The intensity with monochromator is
about 60% of the intensity without monochromator.
-
Mostly a low intensity is caused by the X-ray tube. When you want to be
sure the error is not in the detector you can check the intensity of the
primary beam with a separate detector (for instance a scintillator detector
usable for 8-18 keV).
-
Check the alignment of the goniometer. Perhaps
the primary beam doesn't go through the centre of the goniometer. It is
also possible to test it by using a small 0.4 mm steel ball on the crystal
position and make an image of the primary beam (Always
with absorber and low kV/mA setting)
-
Remove the collimator, switch off X-ray, look into the collimator hole
and check for black powder. Some fast shutters produce black dust
which absorbs the X-ray. Mostly it can be blown away with compressed air.
The black powder can also be on the monochromator crystal. The cause is
when the plunger is too sharp, which then wears the brass. This can be
solved by making the top of the plunger soft soft with a whetstone, not
with a file (mostly the edge, but also no carves on the surface).
-
No images
-
Check in APEX2server Instrument status for warnings that the files cannot
be saved. This can be caused by connection problems with the client PC.
Detector problems
-
Restart: first power up the D8 and wait till it has started up.
Then power on the detector.
-
CCD temperature (can be checked in the image file header)
-
APEX I
-
The default CCD temperature is: -43 - -44 ºC (Water cooler:
+5 ºC). When it goes above -40 ºC the detector should be checked
of leakage and revacuumed. This can only be done in the factory.
-
APEX II
-
The default CCD temperature is -60 - -64 ºC . Because of outgassing
in the detector vacuum this temperature will slowly rise. When it goes
above -50 ºC, also after setting the cooling liquid to -5 ºC,
the detector should be revacuumed. This can only be done by a service engineer.
-
If there is a problem in the image, not related
to reflections, please :
Make two uncalibrated images:
- Start BCP in master mode
- Choose: Tools, test frame
- Make an image: 0.1s, 1 exposure, no shutter, still, no dark&flood,
file name dark0.gadd
- Make the same image with 600s, file name dark600.gadd
- Send both images (in: Program files/BrukerAXS/BCP and BIS/calibrations/)
to the service centre in zipped form (without zipping errors can occur
when transported somewhere with Linux)
-
Other valuable calibration files which service can ask you to send
(in the same directory, #### is the detector number):
-
####_00512_time._dk (time must be the same time the measurements with errors
were done)
-
####_0512_90s._fl
Lytron cooler
-
The default setting is 0 ºC
-
Cooling liquid does not reach 0 ºC (for APEX II). Check if:
-
There is enough clearance (45 cm) for air flow.
-
The condenser coil is obstructed with dust and/or debris
-
The Glycol mix ratio is too high or too low (too much a too little glycol).
This can be caused by condensation inside the cooler which slowsly in time
will dilute the liquid.
-
The filter strainer is clogged, not allowing the coolant to flow freely
through the evaporator ( heat exchanger). Clean
it
-
Check or replace the cooling liquid
If you cleaned or changed the filter inside the Lytron controller be
sure to reinstall its isolation.
Also condensation will occur which drips on electrical connections
below.
So it should not look as: |
 |
-