    Technical support



Safety instructions

X-ray safety

Safety rules as addition to the User Manual:

Beryllium

The detector and X-ray tube contains windows of beryllium. In normal use these area's cannot be reached, but during maintenance, care must be taken not to touch these windows. When by accident the beryllium is touched, wash with soap and water.

Fumes and dust from beryllium metal can be hazardous if inhaled. During use, corrosion of the beryllium window may occur, but it should not be cut, machined or otherwise removed. Tube and detector disposal must comply with the national and local regulations governing beryllium.
 


Maintenance


Cleaning

For cleaning use a damp cloth, moistened with water or isopropyl alcohol. No cleaning agent may enter the goniometer.
 

Tools

Server program

If the server program window is not visible it can be restored by editing a new line with SHOWWINDOW in C:\program files\brukeraxs\bcp and pis\KCCD.SET
 
 

Safety test

This test it limited to some safeties in the goniometer without an internal failure check:

Effic

With Effic the total efficiency is measured from X-ray generation - optics - detector. The measurement below is for PT135 detector and Helios or Montel 200 (both Cu).

  1. Close all programs on the server PC
  2. Start BIS and BCP ( ‘Master (full control)’ option)
  3. Install:
  4. Select in BCP menu ‘Tools’, ‘Half-beam’ and set:
  5. The program shows the intensity in ADU/s (Analog-Digital Units) of the PT135 detector (with Cu-gain of 2.91 ADU/photon)



Troubleshooting

When consulting Service always run D8tools, make a status file (.sts) and send it to the service center


Serial number
 
A service department will often ask for the system serial number. All parts has its own serial  number, but also the whole system has a serial number. It is a number 1-9999 and is normally placed under the table above the keyswitches
However a lot of X8Proteum are delivered without this serial number.

Start-up problems

CCD temperature rising and High pressure cooler with PT135 detector

This is normal during startup. The cooler starts 3 minutes after power on to prevent switching of the fuse because of a too large start current. In this time the pressure (meter on the cooler) will be high and the CCD temperature is slowly rising. But even permanent without cooling the CCD will not rise higher then about 70 C which does not harm the detector. When the cooler starts, pressure will fall down and after a short time also the CCD temperature wil decrease.
 

Shutter does not open


Collision and maximum angle switch

The anti-collision software prevents the goniometer from colliding with the collimator holders or the axis to run past their limits. When a collision takes place because of unexpected circumstances, like a technical error, it is sometimes possible to return:

  1. On the Server computer a menu appears.
  2. Choose the axis and direction to run out of collision.
  3. Press the overrule button at the back of the goniometer. The axis now positions a small step.
  4. If not enough: repeat. As soon as the collision switch is released, the goniometer starts to relocate the zero position.